Pure Data for a Healthy Planet.® YSI Environmental YSI 5200 Recirculating System Monitor Operations Manual
YSI Incorporated 8 5200 Recirculating System Monitor via a digital pager. Used in conjunction with a computer running AquaManager Software, it provi
YSI Incorporated 98 5200 Recirculating System Monitor Day1 Day 2 Day 3 Day 4 Day 5 Day 6 Day 7 Biomass 1000 1020 1040 1061 1082 1104 1126 Feed %
YSI Incorporated 99 5200 Recirculating System Monitor Measured Weight Weigh the amount of food dispensed and enter the value at the Measured Weight
YSI Incorporated 100 5200 Recirculating System Monitor NOTE: Allow enough time to clean the probes or provide tank maintenance without breaching any
YSI Incorporated 101 5200 Recirculating System Monitor 9.3.1 Example—Wiring a Contactor The 5200’s internal relays are not intended to switch heavy
YSI Incorporated 102 5200 Recirculating System Monitor 9.4 Auxiliary System The YSI 5200 has two auxiliary inputs. Use digital inputs for control,
YSI Incorporated 103 5200 Recirculating System Monitor 9.4.1 Input #1 Input #1 is a digital only input, triggered by either a switch open or a switc
YSI Incorporated 104 5200 Recirculating System Monitor User Def.SuffixXXXXMin/Max ValuesScaleSuffixLabelAlarm SystemSet Point Menu Alarm Low Alarm H
YSI Incorporated 105 5200 Recirculating System Monitor Alarm System The alarm system can be turned on or off. When on, an audible alarm will sound o
YSI Incorporated 106 5200 Recirculating System Monitor Section 10. AquaManager Software AquaManager is a 32-bit Windows® application that provides
YSI Incorporated 107 5200 Recirculating System Monitor Section 11. Maintenance It is important to remember that the quality of data obtained will b
YSI Incorporated 9 5200 Recirculating System Monitor Section 3. Installation Installation includes the following sections: Ö Unpacking and Inspect
YSI Incorporated 108 5200 Recirculating System Monitor 11.2 Calibration Checks The sensors are of high quality and should exhibit excellent performa
YSI Incorporated 109 5200 Recirculating System Monitor 10. (5562 Probe only) Activate the 2-point pH calibration protocol from the 5200 display and
YSI Incorporated 110 5200 Recirculating System Monitor 11.4 Sensor Care and Maintenance Once the sensors have been properly installed, remember that
YSI Incorporated 111 5200 Recirculating System Monitor DO Sensor Hex Key Wrench Set Screw Probe Sensor Guard Chemical Cleaning: Remove the membrane
YSI Incorporated 112 5200 Recirculating System Monitor Figure 11-2 5561 Probe Module - DO Sensor Replacement 2. Insert the long end of the hex ke
YSI Incorporated 113 5200 Recirculating System Monitor CAUTION: When using a cotton swab with the YSI 5564 or YSI 5565, be careful NOT to wedge the
YSI Incorporated 114 5200 Recirculating System Monitor 3. After cleaning, check the response and accuracy of the conductivity cell with a calibrati
YSI Incorporated 115 5200 Recirculating System Monitor Section 12. Storage Proper storage between periods of usage will not only extend the life of
YSI Incorporated 116 5200 Recirculating System Monitor rehydrated by soaking the sensor for several hours (overnight is recommended) in a solution w
YSI Incorporated 117 5200 Recirculating System Monitor Section 13. Troubleshooting The following table describes problems you may encounter when us
YSI Incorporated 10 5200 Recirculating System Monitor WARNING: To avoid severe personal injury or damage to the equipment, installation, operation
YSI Incorporated 118 5200 Recirculating System Monitor Problem Possible Solution Sensor Problems Sensor requires cleaning. Follow sensor cleaning p
YSI Incorporated 119 5200 Recirculating System Monitor Range Low In Use (120) Only one parameter can control each relay. In order to use this rel
YSI Incorporated 120 5200 Recirculating System Monitor Section 14. Warranty and Service Information This YSI 5200 Recirculating Monitor ("YSI
YSI Incorporated 121 5200 Recirculating System Monitor U. S. Service Centers Ohio YSI Incorporated • Repair Center • 1725 Brannum Lane • Yellow Spri
YSI Incorporated 122 5200 Recirculating System Monitor
YSI Incorporated 123 5200 Recirculating System Monitor Section 15. System Specifications Temperature Sensor type 10,000 ohm @ 25°C YSI Precision™
YSI Incorporated 124 5200 Recirculating System Monitor Salinity Sensor type calculation from temperature and conductivity Range 0 to 80 ppt Accur
YSI Incorporated 125 5200 Recirculating System Monitor Section 16. Factory Default Settings Setting Factory Default Value Your Value Parameters—
YSI Incorporated 126 5200 Recirculating System Monitor Setting Factory Default Value Your Value Conductivity Alarms No Parameters—DO DO Enabled
YSI Incorporated 127 5200 Recirculating System Monitor Setting Factory Default Value Your Value pH Low Control Relay Disabled pH High Alarm 8.
YSI Incorporated 11 5200 Recirculating System Monitor Note that this is an on-line device that is measuring actual conditions in real time. Composit
YSI Incorporated 128 5200 Recirculating System Monitor Setting Factory Default Value Your Value Feed Timer Start Time 10:00 Feed Timer End Time
YSI Incorporated 129 5200 Recirculating System Monitor Setting Factory Default Value Your Value Baud Rate 19200 Clean Probes Enabled Clean P
YSI Incorporated 130 5200 Recirculating System Monitor Node Log Sheet Node Name Serial Number 52- Software Version Boot Version Ne
YSI Incorporated 131 5200 Recirculating System Monitor Temperature Enabled Yes/No Alarm Yes/No Temperature Scale °F/°C Alarm High Relay Rang
YSI Incorporated 132 5200 Recirculating System Monitor Node Log Sheet Node Name Serial Number 52- Feed Timer Enabled Yes/No Feed Days S
YSI Incorporated 133 5200 Recirculating System Monitor Section 17. International Modem Support The YSI 5201 Modem is FCC Part 68 approved, and UL 1
YSI Incorporated 134 5200 Recirculating System Monitor Section 18. Health and Safety YSI Conductivity solutions: 3161, 3163, 3165, 3167, 3168 and
YSI Incorporated 135 5200 Recirculating System Monitor YSI pH 4.00, 7.00, and 10.00 Buffer Solutions: 3821, 3822, 3823 pH 4 INGREDIENTS: θ P
YSI Incorporated 136 5200 Recirculating System Monitor YSI Zobell Solution: 3682 INGREDIENTS: θ Potassium Chloride θ Potassium Ferrocyanide
YSI Incorporated 137 5200 Recirculating System Monitor YSI Replacement Desiccant 065802 INGREDIENTS θ Calcium Sulfate and Calcium Chloride CA
YSI Incorporated 12 5200 Recirculating System Monitor structures or near a high heat source, AC motor or transformer, radio transmitter or antenna.
YSI Incorporated 138 5200 Recirculating System Monitor Section 19. Required Notice The Federal Communications Commission defines this product as a
YSI Incorporated 139 5200 Recirculating System Monitor Section 20. Accessories Optional Accessories for 5200 Monitor • 5579 Power Supply 12 VDC,
YSI Incorporated 140 5200 Recirculating System Monitor Section 21. Solubility and Pressure/Altitude Tables Table 1: Solubility of Oxygen (mg/L) i
YSI Incorporated 141 5200 Recirculating System Monitor Temp oC Chlorinity 0 Salinity: 0 5.0 ppt 9.0 ppt 10.0 ppt 18.1 ppt 15.0 ppt 27.1 p
YSI Incorporated 142 5200 Recirculating System Monitor Table 2: Calibration Values for Various Atmospheric Pressures and Altitudes PRESSURE ALTIT
YSI Incorporated 143 5200 Recirculating System Monitor Table 3: Conversion Factors for Feet/Meters, Celsius/Fahrenheit, mg/L/ppm TO CONVERT FROM
YSI Incorporated 144 5200 Recirculating System Monitor Section 22. YSI Conductivity Calibration Solution Values Temperature 3161/3167 3163/3168
YSI Incorporated 145 5200 Recirculating System Monitor Section 23. Declaration of Conformity Manufacturer: YSI Incorporated 1725 Brannum Lane
YSI Environmental 1700/1725 Brannum Lane Yellow Springs, OH 45387 USA 800.897.4151 937.767.7241 937.767.9353 fax [email protected]
YSI Incorporated 13 5200 Recirculating System Monitor 3.3.2 Installation Check List 9 Determine optimum mounting location for the probe based on t
YSI Incorporated 14 5200 Recirculating System Monitor The following steps should be followed when wall mounting the 5200 Monitor. 1. Loosely fast
YSI Incorporated 15 5200 Recirculating System Monitor 3.4.2 Optional 6509 – Rail Mount Kit The 5200 Monitor can be easily mounted to pipe or handra
YSI Incorporated 16 5200 Recirculating System Monitor 3.4.4 Optional 6510 – Panel Mount Kit Panel mounting the 5200 Monitor is a simple process usi
YSI Incorporated 17 5200 Recirculating System Monitor 3.5 Installing the 5562 or 5561 Probe Assemblies CAUTION: Keep pH and ORP probe tips IN A MO
YSI Incorporated 18 5200 Recirculating System Monitor For this type of installation, the probe guard is removed and the probe is threaded into a tee
YSI Incorporated 19 5200 Recirculating System Monitor Figure 3-6 Adapter Position The following steps should be followed when mounting the 5205 or
YSI Incorporated 20 5200 Recirculating System Monitor 7. Slide the uncapped end of the PVC pipe up through the 2” u-bolts until there is approximat
YSI Incorporated 21 5200 Recirculating System Monitor Figure 3-9 Placement of Ferrites and Pluggable Sockets 3.6.1 DC Power Input Wiring 12VDC pow
YSI Incorporated 22 5200 Recirculating System Monitor supply” or other power source which is “hazardous live” per UL 3101-1 Section 3.5.2. The 5200
YSI Incorporated 23 5200 Recirculating System Monitor CAUTION: The sensitivity and stability of the monitor will be impaired if the monitor is not
YSI Incorporated 24 5200 Recirculating System Monitor Wiring Diagram for DB-9 to RJ-45 Adapter Signal name (from 5200) RJ-45 pin # Wire color inter
YSI Incorporated 25 5200 Recirculating System Monitor RJ-45 Pin Number RS232 Name Direction RS232 TCP/IP Signal Function 1 DCD I NC P Signals
YSI Incorporated 26 5200 Recirculating System Monitor 3.8 Grounding Information This section contains important installation information regarding g
YSI Incorporated 27 5200 Recirculating System Monitor Section 4. Probe Module Probe Module covers the following: Ö Unpacking the Probe Module Ö
YSI Incorporated i 5200 Recirculating System Monitor Table of Contents SECTION 1. SAFETY ...
YSI Incorporated 28 5200 Recirculating System Monitor 4.3 Features of the YSI 5562 Probe Module Figure 4-2 5562 Probe Module 4.4 Pre
YSI Incorporated 29 5200 Recirculating System Monitor Figure 4-3 Port Plug Removal 3. Locate the port with the connector that corresponds to t
YSI Incorporated 30 5200 Recirculating System Monitor CAUTION: Do not cross thread the sensor nut. Tighten the nut until it is flush with the face
YSI Incorporated 31 5200 Recirculating System Monitor CAUTION: Do not touch the membrane surface. 8. Screw the membrane cap onto the sensor modera
YSI Incorporated 32 5200 Recirculating System Monitor Figure 4-10 Bottom Connectors 2. Holding the cable firmly against the cable connector, turn
YSI Incorporated 33 5200 Recirculating System Monitor Section 5. Operation Operation covers the basic set up and operation of the 5200 Recirculatin
YSI Incorporated 34 5200 Recirculating System Monitor Cond 24.0mS * Sal. 10.4ppt The status LEDs show the current status of each parameter as fo
YSI Incorporated 35 5200 Recirculating System Monitor 5.3 Parameters Menu The Parameters Menu allows you to enter the menu for specific measurement
YSI Incorporated 36 5200 Recirculating System Monitor Figure 5-3 Setpoints and Ranges When these values are entered, the system will operate as ill
YSI Incorporated 37 5200 Recirculating System Monitor Parameter Set Point High Set Point #1 High Set Point #2 Low Set Point #1 Low Set Point #2 H
YSI Incorporated ii 5200 Recirculating System Monitor 7.1 ALARM LOGIC FLOW CHARTS ...
YSI Incorporated 38 5200 Recirculating System Monitor Temp Menu EnabledNOYesControl MenuSet Point MenuSetup MenuAlarm HighAlarm LowSet PointRange Hi
YSI Incorporated 39 5200 Recirculating System Monitor Set Point Menu Enter to Select Press the Enter [ ] key. Enter to Select SetPoint Set Point Me
YSI Incorporated 40 5200 Recirculating System Monitor Range High From the Set Point menu (with the Range High option displayed), press the Enter [
YSI Incorporated 41 5200 Recirculating System Monitor Enter to Select Alarm High NOTE: You may enter the other Temperature System menu options (Cont
YSI Incorporated 42 5200 Recirculating System Monitor Range Low menu selections are Relay Disabled, Relay Internal and Relay SubNet. NOTE: Range Hig
YSI Incorporated 43 5200 Recirculating System Monitor Press the Enter [ ] key to save the changes, or press the [e] or [f] key to select No, then p
YSI Incorporated 44 5200 Recirculating System Monitor Changing the temperature scale automatically recomputes your Set Points. For example, if your
YSI Incorporated 45 5200 Recirculating System Monitor 5.3.3 DO System DO System menu selections are shown below. DO Menu EnabledNOYesControl MenuSet
YSI Incorporated 46 5200 Recirculating System Monitor NOTE: Set Point, Control and Alarm System Menus are only available when a Manual Display Range
YSI Incorporated 47 5200 Recirculating System Monitor Alarm System The alarm system for each measurement parameter can be turned on or off. When on,
YSI Incorporated 3 5200 Recirculating System Monitor Section 1. Safety 1.1 General Safety Information Read all safety information in this manual c
YSI Incorporated 48 5200 Recirculating System Monitor seconds of the powered cycle. The entire cycle is then repeated. Interval mode extends the tim
YSI Incorporated 49 5200 Recirculating System Monitor 5.3.4 pH System pH System menu selections are shown below. pH Menu EnabledNoYesControl MenuSe
YSI Incorporated 50 5200 Recirculating System Monitor Calibration Calibration is covered in Section 6 Calibration. pH can be calibrated at either on
YSI Incorporated 51 5200 Recirculating System Monitor target operating range, it will turn on the pH doser or CO2 injector until the pH returns to t
YSI Incorporated 52 5200 Recirculating System Monitor 5.3.5 ORP System ORP System menu selections are shown below. ORPMenu EnabledNoYesControl MenuS
YSI Incorporated 53 5200 Recirculating System Monitor Calibration Calibration is covered in Section 6 Calibration. See Section 6.2.5 ORP Calibration
YSI Incorporated 54 5200 Recirculating System Monitor 5.3.6 Conductivity System Conductivity System menu selections are shown below. Conductivity Me
YSI Incorporated 55 5200 Recirculating System Monitor NOTE: Set Point, Control and Alarm System Menus are only available when a Manual Display Range
YSI Incorporated 56 5200 Recirculating System Monitor Set the Dose Down to 0 to select normal conductivity control (see Section 5.3.6 Conductivity S
YSI Incorporated 57 5200 Recirculating System Monitor Display Range The 5200’s Conductivity system operates in either Auto ranging or 1 of 4 manual
YSI Incorporated 4 5200 Recirculating System Monitor WARNING: A UL Listed slow-blow fuse with a maximum current rating of 1A must be connected in s
YSI Incorporated 58 5200 Recirculating System Monitor 8. Clean the sensors every 30–90 days, or as needed. A screen prompt on the YSI 5200 display
YSI Incorporated 59 5200 Recirculating System Monitor 9. Press the Enter [ ] key to change the state of the relay to On. R1 R2 R3 R4 Off O
YSI Incorporated 60 5200 Recirculating System Monitor Section 6. Calibration All of the sensors, except temperature, require periodic calibration
YSI Incorporated 61 5200 Recirculating System Monitor Ö If you are using laboratory glassware for calibration, you do not need to remove the probe
YSI Incorporated 62 5200 Recirculating System Monitor 6.2 Calibration Procedures 6.2.1 Accessing the Calibrate Screen 1. From the Run screen, pres
YSI Incorporated 63 5200 Recirculating System Monitor 6.2.2 Dissolved Oxygen Calibration This procedure calibrates dissolved oxygen. Calibrating any
YSI Incorporated 64 5200 Recirculating System Monitor Salinity 30 ppt NOTE: The instrument must be on for at least 20 minutes to polarize the
YSI Incorporated 65 5200 Recirculating System Monitor 2. If your probe module contains a conductivity probe and it is Enabled, skip to Step 3. If y
YSI Incorporated 66 5200 Recirculating System Monitor 6.2.3 Conductivity Calibration This procedure calibrates conductivity and salinity. Note that
YSI Incorporated 67 5200 Recirculating System Monitor NOTE: The sensor must be completely immersed past its vent hole. Using the recommended volumes
YSI Incorporated 5 5200 Recirculating System Monitor Section 2. Introduction Congratulations on your purchase of a sophisticated, yet easy-to-use a
YSI Incorporated 68 5200 Recirculating System Monitor 6.2.4 pH Calibration Calibration1 PointDisplay Actual Value " " Current Temp ValueSc
YSI Incorporated 69 5200 Recirculating System Monitor one-point calibration. The procedure for this calibration is the same as for a 2-point calibra
YSI Incorporated 70 5200 Recirculating System Monitor 12. Press the Enter [ ] key to enter the value. The display will notify you that this is Pass
YSI Incorporated 71 5200 Recirculating System Monitor 6.2.5 ORP Calibration CalibrationDisplay Actual Value " "Current Temp ValueScrolling
YSI Incorporated 72 5200 Recirculating System Monitor -703mv T=72.5°C User Val +307 8. Enter the correct value of the calibration solution you are
YSI Incorporated 73 5200 Recirculating System Monitor Section 7. Alarm/Pager 7.1 Alarm Logic Flow Charts Is Parameteroutside alarmvaluesIs alarm f
YSI Incorporated 74 5200 Recirculating System Monitor EmailIs AutopollerrunningIs Autopoller flagfor this node setIs "Enable EmailAlarm Page Sy
YSI Incorporated 75 5200 Recirculating System Monitor Has power sourcechangedIs Gerneral AlarmenabledIs PagerenabledSendPageYesNoNoYesStopNoYesActiv
YSI Incorporated 76 5200 Recirculating System Monitor Is Auxiliary state =alarm stateIs Alarm holdoff timeset to zeroIs Gerneral AlarmenabledCount
YSI Incorporated 77 5200 Recirculating System Monitor 7.2 General Alarm TemperatureAlarmDO AlarmpH AlarmORP AlarmConductivityAlarmAuxiliary AlarmGen
YSI Incorporated 6 5200 Recirculating System Monitor Front Panel The front panel of the 5200 contains the display and keypad as shown in Figure 2-1
YSI Incorporated 78 5200 Recirculating System Monitor Section 8. Network Multiple 5200 Recirculating Monitors can be connected together via their R
YSI Incorporated 79 5200 Recirculating System Monitor Wire Terminal Connection Solid Blue Wire 1 T (-) White w/Blue Stripe 2 T (+) Solid Oran
YSI Incorporated 80 5200 Recirculating System Monitor 2. Connect the switch to Auxiliary Input #1 or Auxiliary Input #2 using the 3-pole YSI 065942
YSI Incorporated 81 5200 Recirculating System Monitor IN 1 --IN 2 --GND-- IN 1 --IN 2 --GND-- Figure 8-2 Analog Instrument Connection
YSI Incorporated 82 5200 Recirculating System Monitor Section 9. Advanced Setup Advanced Setup includes: Ö System Menu Ö Timers Menu Ö Relay Out
YSI Incorporated 83 5200 Recirculating System Monitor System MenuLocaleData Service PasswordCommunicationsPrint DataData LoggerClear DataCurrent Log
YSI Incorporated 84 5200 Recirculating System Monitor 9.1.1 Service Service menu selections are shown below. Service MenuReset To FactoryClean Probe
YSI Incorporated 85 5200 Recirculating System Monitor Reset After the set number of days has passed and the Clean Probes message appears on the disp
YSI Incorporated 86 5200 Recirculating System Monitor 9.1.2 Communications Communications menu selections are shown below. CommunicationsMenuNetwork
YSI Incorporated 87 5200 Recirculating System Monitor Alarm Logic Flow Charts). After the Alarm Hold Off period expires, the audible alarm (a standa
YSI Incorporated 7 5200 Recirculating System Monitor Status Lights The 6 status lights located on the front panel of the 5200 (see Figure 2-1 Front
YSI Incorporated 88 5200 Recirculating System Monitor Dial Modifier Description , Delay for approx. 1 second W Wait for dial tone The ‘W’
YSI Incorporated 89 5200 Recirculating System Monitor Page Now Page Now sends a test page consisting of the last 3 digits of the serial number of th
YSI Incorporated 90 5200 Recirculating System Monitor 5. After the download is complete, run the file (that you just downloaded) and follow the on
YSI Incorporated 91 5200 Recirculating System Monitor 12. Disconnect power from the 5200. 13. Reapply power to the 5200. 9.1.6 General Alarm See Sec
YSI Incorporated 92 5200 Recirculating System Monitor Log Events The 5200 can Log Events performed, such as a control system, alarm, or timer turnin
YSI Incorporated 93 5200 Recirculating System Monitor 9.2 Timers Menu The Timers Menu includes options for the YSI 5200 to control various timers in
YSI Incorporated 94 5200 Recirculating System Monitor 9.2.1 Feed Timer The Feed Timer System allows control and monitoring of feedings made by the 5
YSI Incorporated 95 5200 Recirculating System Monitor Timed/ContinuousSMTWTFSDaily Weight only available whenFCR is disabledEnabledOz,lb,g,kg001-500
YSI Incorporated 96 5200 Recirculating System Monitor Feeder Prior to operating the feed timer system, the following fields must be configured: Ö F
YSI Incorporated 97 5200 Recirculating System Monitor Select Cont. (Continuous) if feedings are to occur over the entire day. Once a Continuous feed