YSI Incorporated Data Scout Advanced 56
The Configure | User Units (HiLi Xdcr) menu item opens the following window:
Figure 4.42
The two text boxes of this dialog window contain the current values of the Offset and Slope options, respectively, for defining User
Units data read from the transducer. See their ToolTips for more information. Click either text box to edit its original value and then
click OK to change it. These coefficients will be applied to the psi units and are only used if you have specified User Units as the Units
of Pressure for the transducer.
The Configure | Alter Meas. Commands menu item opens the following window:
Figure 4.43
This option affects the response of several SDI-12 primary Pressure measurement commands. If the Press. Only (normal) option is
selected, these transducer commands cause only return a pressure datum. If the Press. and Temp option is selected, these transducer
commands return both a pressure and a temperature measurement datum.